Current Directions in Biomedical Engineering (Sep 2015)

Investigation of the dynamic diameter deformation of vascular stents during fatigue testing with radial loading

  • Boeck Maria,
  • Kamke F.,
  • Grabow N.,
  • Schmidt W.

DOI
https://doi.org/10.1515/cdbme-2015-0124
Journal volume & issue
Vol. 1, no. 1
pp. 519 – 523

Abstract

Read online

Endovascular stents are exposed to cyclic loads resulting from daily activity and pulsatile arterial blood pressure. DIN EN ISO 25539-2 and FDA guideline 1545 recommend durability testing, exposing stents to physiological cyclic loads for a 10 year equivalent. For accelerated testing, the simulated deformation has to be comparable to physiological in-vivo deformation. A new test setup is presented to determine diameter deformation of vascular stents during fatigue testing with radial loading.