JACC: Case Reports (Oct 2019)

A Case of LV Lead Fracture in Submuscular Implant

  • Amit Kumar, MBBS,
  • Vatsal Ladia, MD,
  • Victor Abrich, MD,
  • Luis Scott, MD

Journal volume & issue
Vol. 1, no. 3
pp. 306 – 310

Abstract

Read online

Lead fracture is one of the mechanical complications of implanted cardiac devices. We present a case of mechanical stress-induced left ventricular lead fracture attributed to the hypermobility of the device secondary to sub-muscular implantation of the device without anchoring it. (Level of Difficulty: Intermediate.)

Keywords