Frontiers in Veterinary Science (Jan 2024)

Visualization of anatomical structures in the fetlock region of the horse using cone beam computed tomography in comparison with conventional multidetector computed tomography

  • Jonathan Bierau,
  • Antonio M. Cruz,
  • Christoph Koch,
  • Gabriel Manso-Diaz,
  • Kathrin Büttner,
  • Carsten Staszyk,
  • Michael Röcken

DOI
https://doi.org/10.3389/fvets.2023.1278148
Journal volume & issue
Vol. 10

Abstract

Read online

IntroductionCone-beam computed tomography (CBCT) is regarded as a convenient and suitable alternative to conventional computed tomography. However, in the horse, the quality of obtained data sets needs to be evaluated. Therefore, the aim of this study was to compare the visibility and accessibility of clinically relevant anatomical structures displayed in CBCT and conventional multidetector computed tomography (MDCT).Materials and methodsTwenty-nine limbs from horses euthanized for reasons unrelated to this study were used. Native and intraarticular contrast scans of the fetlock (CBCT vs. MDCT) were performed. The visibility and accessibility of selected anatomical structures were blindly scored by three independent experienced observers using a scoring system previously reported and adapted to the fetlock joint.ResultsOnly minor differences between CBCT and MDCT were identified concerning the diagnostic quality of images for osseous structures. Soft tissue structures were better evaluated on MDCT images. In CBCT as well as in MDCT articular cartilage could only be visualized after intraarticular injection of contrast medium.Discussion/conclusionCone beam computed tomography of the fetlock is a useful and reliable diagnostic tool when evaluating osseous structures and delineating articular cartilage with contrast medium. However, this modality is limited for assessing soft tissues structures.

Keywords