Zhongguo yiliao qixie zazhi (Nov 2024)

Development of Non-Invasive Continuous Blood Pressure Measurement System Based on Multiple Feature Parameters

  • Mingyue LI,
  • Zhiying YUAN,
  • Ruowei LI,
  • Jieying SHAN,
  • Jilun YE,
  • Xu ZHANG,
  • Hui YU,
  • Xiaoyu WU

DOI
https://doi.org/10.12455/j.issn.1671-7104.230417
Journal volume & issue
Vol. 48, no. 6
pp. 670 – 677

Abstract

Read online

To address the issue of the difficulty in implementing non-invasive continuous blood pressure measurement technology in China, this study developed a high-performance synchronous electrocardiogram (ECG) and photoplethysmography (PPG) signal acquisition system. A PC-based human-computer interaction software platform was constructed, and continuous blood pressure measurement-related algorithms were integrated. Multiple feature parameters such as pulse wave transit time based on synchronous ECG and PPG signals were extracted, enabling non-invasive continuous blood pressure measurement. To verify the measurement accuracy of the system, tests and comparative verifications were carried out. The results demonstrated that the system could meet the requirements of relevant standards and have good application value.