BMC Research Notes (Jan 2018)

Intrarater reliability of the Humac NORM isokinetic dynamometer for strength measurements of the knee and shoulder muscles

  • Bas Habets,
  • J. Bart Staal,
  • Marsha Tijssen,
  • Robert van Cingel

DOI
https://doi.org/10.1186/s13104-018-3128-9
Journal volume & issue
Vol. 11, no. 1
pp. 1 – 5

Abstract

Read online

Abstract Objective To determine the intrarater reliability of the Humac NORM isokinetic dynamometer for concentric and eccentric strength tests of knee and shoulder muscles. Results 54 participants (50% female, average age 20.9 ± 3.1 years) performed concentric and eccentric strength measures of the knee extensors and flexors, and the shoulder internal and external rotators on two different Humac NORM isokinetic dynamometers, which were situated at two different centers. The knee extensors and flexors were tested concentrically at 60° and 180°/s, and eccentrically at 60° s. Concentric strength of the shoulder internal and external rotators, and eccentric strength of the external rotators were measured at 60° and 120°/s. We calculated intraclass correlation coefficients (ICCs), standard error of measurement, standard error of measurement expressed as a %, and the smallest detectable change to determine reliability and measurement error. ICCs for the knee tests ranged from 0.74 to 0.89, whereas ICC values for the shoulder tests ranged from 0.72 to 0.94. Measurement error was highest for the concentric test of the knee extensors and lowest for the concentric test of shoulder external rotators.

Keywords