Sensors (Sep 2024)

New, Optimized Skin Calorimeter Version for Measuring Thermal Responses of Localized Skin Areas during Physical Activity

  • Miriam Rodríguez de Rivera,
  • Pedro Jesús Rodríguez de Rivera

DOI
https://doi.org/10.3390/s24185927
Journal volume & issue
Vol. 24, no. 18
p. 5927

Abstract

Read online

We present an optimized version of the skin calorimeter for measuring localized skin thermal responses during physical activity. Enhancements include a new holding system, more sensitive thermopiles, and an upgraded spiked heat sink for improved efficiency. In addition, we used a new, improved calorimetric model that takes into account all the variables that influence the measurement process. Resolution in power measurement is 1 mW. Performance tests under air currents and movement disturbances showed that the device maintains high accuracy; the deviation produced by these significant disturbances is less than 5%. Human subject tests, both at rest and during exercise, confirmed its ability to accurately measure localized skin heat flux, heat capacity, and thermal resistance (less than 5% uncertainty). These findings highlight the calorimeter’s potential for applications in sports medicine and physiological studies.

Keywords