BMC Anesthesiology (Feb 2024)

Feasibility and accuracy of noninvasive continuous hemoglobin monitoring using transesophageal photoplethysmography in porcine model

  • Ling Peng,
  • Long Zhao,
  • Xue Zhang,
  • Yi Zhang,
  • Meng Ding,
  • Zhibin Lin,
  • Hao Jiang,
  • Yuchen Huang,
  • Bo Gao,
  • Wei Wei

DOI
https://doi.org/10.1186/s12871-024-02435-7
Journal volume & issue
Vol. 24, no. 1
pp. 1 – 9

Abstract

Read online

Abstract Background Continuous and noninvasive hemoglobin (Hb) monitoring during surgery is essential for anesthesiologists to make transfusions decisions. The aim of this study was to investigate the feasibility and accuracy of noninvasive and continuous Hb monitoring using transesophageal descending aortic photoplethysmography (dPPG) in porcine model. Methods Nineteen landrace pigs, aged 3 to 5 months and weighing 30 to 50 kg, were enrolled in this study. A homemade oximetry sensor, including red (660 nm) and infrared (940 nm) lights, was placed in the esophagus for dPPG signal detection to pair with the corresponding reference Hb values (Hbi−STAT) measured by blood gas analysis. The decrease and increase changes in Hb concentration were achieved by hemodilution and transfusion. Metrics, including alternating current (AC), direct current (DC), and AC/DC for both red and infrared light were extracted from the dPPG signal. A receiver operating characteristic (ROC) curve was built to evaluate the performance of dPPG metrics in predicting the Hb “trigger threshold” of transfusion (Hb 100 g/L). Agreement and trending ability between Hb measured by dPPG (HbdPPG) and by blood gas analysis were analyzed by Bland-Altman method and polar plot graph. Error grid analysis was also performed to evaluate clinical significance of HbdPPG measurement. Results The dPPG signal was successfully detected in all of the enrolled experimental pigs, without the occurrence of a continuous loss of dPPG signal for 2 min during the entire measurement. A total of 376 pairs of dPPG signal and Hbi−STAT were acquired. ACred/DCred and ACinf/DCinf had moderate correlations with Hbi−STAT, and the correlation coefficients were 0.790 and 0.782, respectively. The areas under the ROC curve for ACred/DCred and ACinf/DCinf in predicting Hbi−STAT 100 g/L were 0.90 and 0.83, respectively. Bland-Altman analysis and polar plot showed a small bias (1.69 g/L) but a wide limit of agreement (-26.02–29.40 g/L) and a poor trend ability between HbdPPG and Hbi−STAT. Clinical significance analysis showed that 82% of the data lay within the Zone A, 18% within the Zone B, and 0% within the Zone C. Conclusion It is feasible to establish a noninvasive and continuous Hb monitoring by transesophageal dPPG signal. The ACred/DCred extracted from the dPPG signal could provide a sensitive prediction of the Hb threshold for transfusion. The Hb concentration measured by dPPG signal has a moderate correlation with that measured by blood gas analysis. This animal study may provide an experimental basis for the development of bedside HbdPPG monitoring in the future.

Keywords