Frontiers in Cardiovascular Medicine (Jun 2022)

Bioprosthetic Valve Fracturing: In vitro Testing of Edwards PERIMOUNT Model P 2900

  • Hendrik Ruge,
  • Hendrik Ruge,
  • Hector A. Alvarez-Covarrubias,
  • Hector A. Alvarez-Covarrubias,
  • Oliver Deutsch,
  • Oliver Deutsch,
  • Zahra Alalawi,
  • Keti Vitanova,
  • Keti Vitanova,
  • Rüdiger Lange,
  • Rüdiger Lange

DOI
https://doi.org/10.3389/fcvm.2022.859088
Journal volume & issue
Vol. 9

Abstract

Read online

BackgroundBioprosthetic valve fracturing (BVF) results in low gradients following valve-in-valve transcatheter aortic valve replacement (ViV-TAVR). For the commonly used Edwards PERIMOUNT valve data from bench-testing are lacking to provide technical specifications for successful BVF during ViV-TAVR.MethodsUsing four Perimount 19- and 21-mm valves, in-vitro high-pressure balloon valvuloplasty with the True Dilatation Balloon Valvuloplasty Catheter and Atlas Gold PTA Dilatation Catheter was performed to analyze balloon-oversizing and pressure-thresholds to successfully achieve BVF.ResultsHigh-pressure balloons one millimeter larger than the labeled valve size and pressure rates of 20 atm (for Perimount 19-mm) and > 22 atm (for Perimount 21-mm) were required to achieve BVF. Caliper measurements demonstrated 2.5 mm (Perimount 19-mm) and 1.5 mm (Perimount 21-mm) enlarged inner prosthetic diameters after BVF. The Atlas TM Gold PTA Dilatation Catheter achieved BVF with the Perimount 21-mm, whereas the True TM Dilatation Balloon Valvuloplasty Catheter failed in the Perimount 21-mm either for balloon-rupture or pinhole-defect.ConclusionBoth 19-mm and 21-mm Perimount P 2900 are amendable to BVF, thereby increasing the inner prosthetic diameter. High-pressure balloons 1 mm larger than the labeled valves are essential for this purpose, and the Atlas Gold PTA Dilatation Catheter alone should ensure success in the 21-mm prosthetics.

Keywords