BMC Surgery (Jul 2023)

An experimental study on the impact of prosthesis temperature on the biomechanical properties of bone cement fixation

  • Wanzhuo Chen,
  • Haining Zhang

DOI
https://doi.org/10.1186/s12893-023-02079-3
Journal volume & issue
Vol. 23, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Purpose To investigate the effect of the femoral component and tibial plateau component temperature on the strength of cement fixation during total knee arthroplasty (TKA). Methods Femoral prosthesis, tibial plateau prosthesis, and polypropylene mold base were used to simulate TKA for bone cement fixation. Pre-cooling or pre-warming of femoral and tibial plateau components at different temperatures (4 °C, 15 °C, 25 °C, 37 °C, 45 °C), followed by mixing and stirring of bone cement at laboratory room temperature (22 °C), were performed during research. The prosthesis and the base adhered together, and the bone cement was solidified for 24 h at a constant temperature of 37 °C to verify the hardness of the bone cement with a push-out test. Results The push-out force of the femoral prosthesis after fixation was higher than that of the tibial plateau prosthesis, and with the increase of the prosthesis temperature, the push-out force after fixation of the bone cement also increased linearly and the porosity of the prosthetic cement in the tibia and femur decreased as the temperature increased. Conclusion Without changing the mixing temperature and solidification temperature, the fixation strength of the femoral prosthesis is higher than that of the tibial plateau prosthesis. Properly increasing the temperature of the prosthesis can increase the push-out force of the fixation strength.

Keywords