Timepix3: Compensation of Thermal Distortion of Energy Measurement

  • Martin Urban,
  • Ondrej Nentvich,
  • Lukas Marek,
  • David Hladik,
  • Rene Hudec,
  • Ladislav Sieger

DOI
https://doi.org/10.3390/s23063362
Journal volume & issue
Vol. 23, no. 6
p. 3362

Abstract

Read online

The Timepix3 is a hybrid pixellated radiation detector consisting of a 256 px × 256 px radiation-sensitive matrix. Research has shown that it is susceptible to energy spectrum distortion due to temperature variations. This can lead to a relative measurement error of up to 35% in the tested temperature range of 10 °C to 70 °C. To overcome this issue, this study proposes a complex compensation method to reduce the error to less than 1%. The compensation method was tested with different radiation sources, focusing on energy peaks up to 100 keV. The results of the study showed that a general model for temperature distortion compensation could be established, where the error in the X-ray fluorescence spectrum of Lead (74.97 keV) was reduced from 22% to less than 2% for 60 °C after the correction was applied. The validity of the model was also verified at temperatures below 0 °C, where the relative measurement error for the Tin peak (25.27 keV) was reduced from 11.4% to 2.1% at −40 °C. The results of this study demonstrate the effectiveness of the proposed compensation method and models in significantly improving the accuracy of energy measurements. This has implications for various fields of research and industry that require accurate radiation energy measurements and cannot afford to use power for cooling or temperature stabilisation of the detector.