Current Directions in Biomedical Engineering (Sep 2023)

Immediate stabilization of pedicle screws

  • Wild Michael de,
  • Zimmermann Simon,
  • Klein Karina,
  • Steffen Thomas,
  • Schlottig Falko,
  • Hasler Carol,
  • Rechenberg Brigitte von

DOI
https://doi.org/10.1515/cdbme-2023-1004
Journal volume & issue
Vol. 9, no. 1
pp. 13 – 16

Abstract

Read online

This study was designed as proof of principle and safety test of the novel technique, the Immediate Stabilization System (ISS). The technique is designed to immediately stabilize polymer-augmented pedicle screws (PAS) in deficient bone and avoid complications of loosening pedicle screws at the bone-screw interface, especially in osteoporotic patients. A polymer sleeve was designed as augmentation to improve screw anchorage after drilling the screw hole. By applying ultrasonic energy, the polymeric tube was molded into the pores of the host bone forming a strong and uniform bond with the adjacent bone. The original screw was then implanted into the denser bony environment leading to an enhanced immediate stability. The ISS-treated implants were compared to conventionally placed pedicle screws in ex-vivo cadaver bones (2 sheep spines, n = 6 implants per spine, total 12 screws) and in-vivo in a spinal sheep model (Swiss alpine sheep, n = 5, 4 implants per animal, total 20 screws). The primary stability of ISS-treated pedicle screws was increased in ex-vivo bone (+24% insertion torque (IT)) and in-vivo (+32.9% IT) in sheep spine. Removal torque (RT) was lower in the in PAS tested for 8 weeks in-vivo. The ISS technology demonstrated improved anchorage of pedicle screws in ex-vivo cadaver bones as well as in-vivo studies in sheep spine.

Keywords