PLoS ONE (Jan 2021)

Pressure gradient measurement to verify hemodynamic results of the chimney endovascular aortic repair (chEVAR) technique.

  • Artur Igor Milnerowicz,
  • Aleksandra Milnerowicz,
  • Tomasz Bańkowski,
  • Marcin Protasiewicz

DOI
https://doi.org/10.1371/journal.pone.0249549
Journal volume & issue
Vol. 16, no. 4
p. e0249549

Abstract

Read online

PurposeThe use of the pressure gradient measurements to assess the renal artery flow hemodynamics after chimney endovascular aortic repair (chEVAR).MethodsThe study was a prospective analysis of 37 chEVAR procedures performend in 24 patients with perirenal aortic aneurysm. In all patients the measurement of: distal renal artery pressure (Pd), aortic pressure (Pa), Pd/Pa ratio (Pd/Pa) and mean gradient (MG) between the aorta and the distal renal artery were performed. Measurements were taken with 0.014 inch pressure wire catheter before and after the chEVAR procedure. MG greater than 9 mmHg and Pd/Pa ratio below 0.90 were considered as the measures of a significant decrease in distal pressure that limited flow in renal arteries. The 6 month follow-up computed tomographic angiography (CTA) was performed in all patients to diagnose potential endoleak presence and to verify the patency of the chimney stent-grafts.ResultsAll procedures were successful, and no periprocedural complications were observed in any of the patients. The mean gradient values before and after the chimney implantation did not change significantly (6,2±2,0 mmHg and 6,8±2,2 mmHg, respectively). Similarly, no significant change in Pd/Pa values was noted with the value of 0.9 observed both before and after the procedure. All chimney stents were patent on the control CTA. Type Ia endoleak was found in 4 (10.8%) patients.ConclusionsThe application of the described technique seems to be a safe method which allows a direct measurement of renal artery flow hemodynamics before and after chimney implantation during the chEVAR technique. The use of covered balloon expandable stents, ensures the proper blood flow in the renal arteries during the chEVAR technique.