A Highly Sensitive Dual-Drive Microfluidic Device for Multiplexed Detection of Respiratory Virus Antigens

  • Xiaohui Yang,
  • Yixian Li,
  • Josh Zixi Lee,
  • Yuanmin Sun,
  • Xin Tan,
  • Yijie Liu,
  • Yang Yu,
  • Huiqiang Li,
  • Xue Li

DOI
https://doi.org/10.3390/mi15060685
Journal volume & issue
Vol. 15, no. 6
p. 685

Abstract

Read online

Conventional microfluidic systems that rely on capillary force have a fixed structure and limited sensitivity, which cannot meet the demands of clinical applications. Herein, we propose a dual-drive microfluidic device for sensitive and flexible detection of multiple pathogenic microorganisms antigens/antibodies. The device comprises a portable microfluidic analyzer and a dual-drive microfluidic chip. Along with capillary force, a second active driving force is provided by a removable self-driving valve in the waste chamber. The interval between these two driving forces can be adjusted to control the reaction time in the microchannel, optimizing the formation of antigen-antibody complexes and enhancing sensitivity. Moreover, the material used in the self-driving valve can be changed to adjust the active force strength needed for different tests. The device offers quantitative analysis for respiratory syncytial virus antigen and SARS-CoV-2 antigen using a 35 μL sample, delivering results within 5 min. The detection limits of the system were 1.121 ng/mL and 0.447 ng/mL for respiratory syncytial virus recombinant fusion protein and SARS-CoV-2 recombinant nucleoprotein, respectively. Although the dual-drive microfluidic device has been used for immunoassay for respiratory syncytial virus and SARS-CoV-2 in this study, it can be easily adapted to other immunoassay applications by changing the critical reagents.

Keywords