Investigation of balloon dilation devices for treatment of Eustachian tube dysfunction

  • Schubert Julia,
  • Wilfling Tamara,
  • Paasche Gerrit,
  • Lenarz Thomas,
  • Schümann Kerstin,
  • Grabow Niels,
  • Schmidt Wolfram,
  • Schmitz Klaus-Peter

DOI
https://doi.org/10.1515/cdbme-2018-0127
Journal volume & issue
Vol. 4, no. 1
pp. 529 – 533

Abstract

Read online

The presented three different balloon dilation devices for treating Eustachian tube dysfunction were compared regarding their geometries and designs. The balloon dilation behaviour was investigated by an in vitro test setup consisting of a test chamber with 37 °C water, a 2-axis laser scanner and a pressure controller. All balloons could be properly dilated up to the given nominal pressure (NP) or to rated burst pressure (RBP). The balloons reached their expected balloon diameter and length. The compliance data were 1.07 %/atm (TubaVent short), 1.16 %/atm (Acclarent Aera) and 1.28 %/atm (XprESS LoProfile). The measured profile and compliance data can be used for development of new devices for balloon dilation Eustachian tuboplasty.