Journal of Arrhythmia (Jun 2019)

Inappropriate implantable cardioverter defibrillator shocks due to atrial far‐field on the tip‐to‐ring channel for lead dislodgement

  • Sonia Ferretto,
  • Katia Brunzin,
  • Katiuscia Pettenuzzo,
  • René Nangah Suh,
  • Cristina Canali,
  • Francesco Di Pede

DOI
https://doi.org/10.1002/joa3.12180
Journal volume & issue
Vol. 35, no. 3
pp. 562 – 564

Abstract

Read online

Abstract We reported a case of inappropriate implantable cardioverter defibrillator shocks, due to atrial far‐field on the tip‐to‐ring channel of the fast electrical activity during atrial fibrillation, caused by lead dislocation in the right ventricle outflow tract. During these episodes the can‐to‐right ventricle coil signal correctly recorded the ventricular activity. The shock storm stopped when an antitachycardia pacing restored sinus rhythm.