Greater efficiency observed 12 months post-implementation of an automatic tube sorting and registration system in a core laboratory

  • Ucar Fatma,
  • Erden Gonul,
  • Yavuz-Taslipinar Mine,
  • Ozturk Gulfer,
  • Ginis Zeynep,
  • Bulut Erdem,
  • Delibas Namik

Journal volume & issue
Vol. 35, no. 1
pp. 1 – 6

Abstract

Read online

Background: Sample classification and registration have been recognized as important and time-consuming processes in laboratories. There is increasing pressure on laboratories to automate processes due to intense workload and reduce manual procedures and errors. The aim of the present study was to evaluate the positive effects of an automatic tube registration and sorting system on specimen processing. Methods: An automatic tube registration and sorting system (H CTS2000 M K2, m-u-t AG, Wedel, Germany) was evaluated. Turnaround time (TAT), rate of sample rejection and unrealized tests were examined 12 months preand post-implementation of the automatic tube sorting and registration system. Results: The mean TAT of routine chemistry immunoassay, complete blood cell count (CBC) and coagulation samples were significantly improved (P 0.05). Conclusions: By reducing delays and errors in the preanalytical processing and sorting of samples, significant improvements in specimen processing were observed after im - plementation of the system. These results suggest that an automatic tube registration and sorting system may also be used to improve specimen processing in a higher-volume core laboratory.