Agreement between Two Devices for Measuring Pupil Diameter in Patients Implanted with Multifocal Intraocular Lenses

  • Joaquín Fernández,
  • Noemí Burguera,
  • Carlos Rocha-de-Lossada,
  • Marina Rodríguez-Calvo-de-Mora,
  • Manuel Rodríguez-Vallejo

DOI
https://doi.org/10.3390/vision7020040
Journal volume & issue
Vol. 7, no. 2
p. 40

Abstract

Read online

The purpose of this study was to evaluate the agreement between three methods for measuring pupil size in patients implanted with multifocal intraocular lenses (MIOLs): Keratograph 5M (K5M), Pentacam AXL Wave (PW), and a simple hand ruler. Sixty-nine subjects implanted with MIOLs and measured at the three-month follow-up visit were included in this retrospective analysis. K5M and PW were used to measure the photopic (PP) and mesopic (MP) pupil sizes, and a hand ruler was used to measure the pupil under environmental light conditions (135 lux). The Bland–Altman method with its limits (LoAs) was used to assess the agreement. The median PP was 2.8, 2.95, and 3 mm for K5M, PW, and the ruler, respectively (p p p = 0.44). The LoAs for the difference in PP between K5M and PW was 0.63 mm. The mean difference for MP between K5M and PW was 0.04 mm (p = 0.34) with LoAs of 0.72 mm. MP measured with K5M and PW could be considered interchangeable, although a correction of −0.3 mm (IC95%: −0.23 to −0.39) should be applied to PP measured with PW to attain the K5M mean.