Jornal Brasileiro de Pneumologia (Sep 2013)

Evaluation of manual resuscitators used in ICUs in Brazil

  • Tatiana de Arruda Ortiz,
  • Germano Forti Junior,
  • Marcia Souza Volpe,
  • Marcelo do Amaral Beraldo,
  • Marcelo Britto Passos Amato,
  • Carlos Roberto Ribeiro Carvalho,
  • Mauro Roberto Tucci

DOI
https://doi.org/10.1590/S1806-37132013000500010
Journal volume & issue
Vol. 39, no. 5
pp. 595 – 603

Abstract

Read online

OBJECTIVE: To evaluate the performance of manual resuscitators (MRs) used in Brazil in accordance with international standards. METHODS: Using a respiratory system simulator, four volunteer physiotherapists employed eight MRs (five produced in Brazil and three produced abroad), which were tested for inspiratory and expiratory resistance of the patient valve; functioning of the pressure-limiting valve; and tidal volume (VT) generated when the one-handed and two-handed techniques were used. The tests were performed and analyzed in accordance with the American Society for Testing and Materials (ASTM) F920-93 criteria. RESULTS: Expiratory resistance was greater than 6 cmH2O . L−1 . s−1 in only one MR. The pressure-limiting valve, a feature of five of the MRs, opened at low pressures (< 17 cmH2O), and the maximal pressure was 32.0-55.9 cmH2O. Mean VT varied greatly among the MRs tested. The mean VT values generated with the one-handed technique were lower than the 600 mL recommended by the ASTM. In the situations studied, mean VT was generally lower from the Brazilian-made MRs that had a pressure-limiting valve. CONCLUSIONS: The resistances imposed by the patient valve met the ASTM criteria in all but one of the MRs tested. The pressure-limiting valves of the Brazilian-made MRs usually opened at low pressures, providing lower VT values in the situations studied, especially when the one-handed technique was used, suggesting that both hands should be used and that the pressure-limiting valve should be closed whenever possible.

Keywords