Tehnički Vjesnik (Jan 2017)

Calibration procedure and uncertainty analysis of an electronic scale based on two-current source supplied circuit

  • Wojciech Walendziuk,
  • Adam Idzkowski

DOI
https://doi.org/10.17559/TV-20141026162349
Journal volume & issue
Vol. 24, no. Supplement 1
pp. 93 – 97

Abstract

Read online

The article presents the prototype version of an original weight scale in which a non-conventional supplying circuit of strain gauge load cells is applied. The paper discusses construction of an electrical circuit, the way of strain gauge transducer connection and the measuring procedure. As far as the authors know, this novel electrical circuit and its application in a weight scale system have never been mentioned in literature. The specifications for direct-reading weight scales typically include readability, repeatability, linearity and eccentricity. The values of these parameters were estimated and used for calculation of the weight measurement uncertainty in the tested measuring range of the designed device.

Keywords