He jishu (Jan 2024)

Simulation of internal electric field of particle source water absorbed dose absolute measurement device

  • HANG Zhongbin,
  • LIU Yuntao,
  • SONG Mingzhe,
  • WEI Kexin,
  • WANG Hongyu,
  • LIU Chuanfeng,
  • TENG Zhongbin,
  • GENG Xuan

DOI
https://doi.org/10.11889/j.0253-3219.2024.hjs.47.010404
Journal volume & issue
Vol. 47, no. 1
pp. 95 – 103

Abstract

Read online

BackgroundThe physical quantity concerned in the clinical application of 125I particle source brachytherapy is the 1 cm water absorbed dose rate D˙w,1 cm. However, there is no corresponding standard for this physical quantity in China. It is planned to develop an ionization chamber for the absolute measurement of the water absorbed dose rate of the 125I particle source as a standard device, hence the internal electric field of the ionization chamber must first be analyzed to obtain the most reasonable design scheme to satisfy the condition that the electric field intensity is uniformly distributed in the ionization chamber.PurposeThis study aims to simulate the internal electric field of the ionization chamber for the design of particle source water absorbed dose absolute measurement device.MethodsFirstly, the internal model of the ionization chamber was established by Maxwell software to simulate the distribution of electric field intensity under the six variables of the ionization chamber: with or without a protective electrode, different protective electrode ring width, different insulation ring width, different grid number, different grid shape and different grid thickness. Then, qualitative and quantitative analysis were carried out using finite element method. Finally, the influence of different variables on the distribution of electric field intensity in the ionization chamber of the absolute measurement of water absorbed dose of 125I particle source was obtained.ResultsAnalysis results show the ratio of the width of the guard electrode to the radius of the collector electrode must be not less than 2 for design of the ionization chamber. The edge effect at the edge of the collector is increased with width increment of the insulator ring, hence the width of the insulator ring should be reduced as much as possible. When the number of grid electrodes is 15, the variation of the electric field intensity can be reduced to about 1%. When the grid section is rectangular, the variation in electric field intensity is small compared with the circular and triangular grid sections. The larger the thickness of the grid, the more severe the edge effect at the edge of the grid, and the thickness of the grid should be reduced as much as possible.ConclusionsThe uniformity of the electric field can be effectively improved by increasing the number of grid electrodes. Results of this study is helpful to optimal design of the standard device for measuring the absolute water absorbed dose of 125I particle source.

Keywords