PLoS ONE (Jan 2019)

Intra-arterial pulse wave analysis during thrombectomy for the assessment of collateral status - A feasibility study.

  • Johanna Sandmann,
  • Thorsten Sichtermann,
  • Franziska Sonja Müschenich,
  • Hadi Nasri,
  • Sarah Heringer,
  • Anastasios Mpotsaris,
  • Martin Kramer,
  • Martin Wiesmann,
  • Omid Nikoubashman

DOI
https://doi.org/10.1371/journal.pone.0210572
Journal volume & issue
Vol. 14, no. 1
p. e0210572

Abstract

Read online

PurposeKnowledge of the collateralization of an occluded vessel is important for the risk-benefit analysis of difficult revascularization maneuvers during mechanical thrombectomy. If the territory behind a clot is well perfused, one could desist from performing a risky thrombectomy maneuver. The arterial pulse pressure curve may serve as an indicator for the collateralization of an occluded target vessel. We investigated the feasibility of arterial pulse measurements with a standard microcatheter.MethodsWe measured the intra-arterial blood pressure proximal and distal to the clot in 40 thrombectomy maneuvers in a porcine stroke model. We used a microcatheter (Trevo Pro 18, Stryker, Kalamazoo, CA, USA), a pressure transducer (MEMSCAP SP844), an AdInstruments Powerlab 16/35 workstation, and LabChart 8 Software (AdInstruments, Dunedin, New Zealand).ResultsMedian arterial blood pressure proximal and distal to the clot was 96.0 mmHg (IQR, 23.8 mmHg) and 47.5 mmHg (IQR, 43.5 mmHg), respectively (p ConclusionIt is possible to measure the pulse pressure proximal and distal to the clot with a standard microcatheter used during mechanical thrombectomy.