Effect of lighting conditions on implantable collamer lens vault: Influence of anterior chamber and lens parameters
Zhenhao Song,
Qi Li,
Ying Xiong,
Yingyan Mao,
Xiaofei Wang
Affiliations
Zhenhao Song
Key Laboratory for Biomechanics and Mechanobiology of Ministry of Education, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medical Engineering, Beihang University, Beijing, China
Qi Li
Key Laboratory for Biomechanics and Mechanobiology of Ministry of Education, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medical Engineering, Beihang University, Beijing, China
Ying Xiong
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing, 100730, China
Yingyan Mao
Beijing Institute of Ophthalmology, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing, 100730, China; Beijing Advanced Innovation Center for Big Data-Based Precision Medicine, Beihang University & Capital Medical University, Beijing Tongren Hospital, Beijing, 100730, China; Corresponding author. Beijing Institute of Ophthalmology, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing, 100730, China
Xiaofei Wang
Key Laboratory for Biomechanics and Mechanobiology of Ministry of Education, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medical Engineering, Beihang University, Beijing, China; Eye Hospital, School of Ophthalmology and Optometry and School of Biomedical Engineering, Wenzhou Medical University, Wenzhou, China; Corresponding author. School of Biological Science and Medical Engineering Beihang University Room 424, Building 5, 37 Xueyuan Road, Beijing, 100191, China.
Purpose: To investigate the change in the vault of the implantable collamer lens (ICL) under dark-to-light conditions and its association with anterior chamber and lens parameters in patients undergoing ICL surgery. Methods: In 76 eyes from 40 patients, preoperative anterior chamber volume (ACV), pupil diameter (PD), anterior chamber angle, central corneal thickness (CCT), white-to-white (WTW), lens thickness (LT), axial length (AL), spherical equivalent (SE) and patient's age were collected. Postoperative vault, PD and LT were measured under dark and light conditions using anterior segment optical coherence tomography (CASIA2; TOMEY, Japan), and changes and lens displacement under dark-to-light conditions were calculated. Mixed-effects models were used to analyze the correlation between the vault change and the anterior chamber and lens parameters of all subjects and the high-vault subgroup. Results: The vault under light condition (648.36 ± 304.47 μm) was significantly smaller compared to the vault under dark condition (708.89 ± 316.15 μm). In all patients, vault change increased with the increase of age, lens displacement and PD change; and increased with the decrease of ACV, LT change and baseline vault (under dark condition). In the high-vault subgroup, vault change increased with the increase of CCT, lens displacement and PD change; and increased with the decrease of ACV. Conclusions: ICL vault changes significantly from dark to light, influenced by age, ACV, PD change, LT change, lens displacement, and baseline vault. A higher baseline vault is correlated with a larger LT change, affecting the levels of accommodation under dark-to-light transition.