International Journal of Neonatal Screening (Dec 2017)

Extended and Fully Automated Newborn Screening Method for Mass Spectrometry Detection

  • Stefan Gaugler,
  • Jana Rykl,
  • Irene Wegner,
  • Tamara von Däniken,
  • Ralph Fingerhut,
  • Götz Schlotterbeck

DOI
https://doi.org/10.3390/ijns4010002
Journal volume & issue
Vol. 4, no. 1
p. 2

Abstract

Read online

A new and fully automated newborn screening method for mass spectrometry was introduced in this paper. Pathological relevant amino acids, acylcarnitines, and certain steroids are detected within 4 min per sample. Each sample is treated in an automated and standardized workflow, where a mixture of deuterated internal standards is sprayed onto the sample before extraction. All compounds showed good linearity, and intra- and inter-day variation lies within the acceptance criteria (except for aspartic acid). The described workflow decreases analysis cost and labor while improving the sample traceability towards good laboratory practice.

Keywords