Photoacoustics (Apr 2024)

Mid-infrared all-fiber light-induced thermoelastic spectroscopy sensor based on hollow-core anti-resonant fiber

  • Weipeng Chen,
  • Shunda Qiao,
  • Ying He,
  • Jie Zhu,
  • Kang Wang,
  • Lei Qi,
  • Sheng Zhou,
  • Limin Xiao,
  • Yufei Ma

Journal volume & issue
Vol. 36
p. 100594

Abstract

Read online

In this article, a mid-infrared all-fiber light-induced thermoelastic spectroscopy (LITES) sensor based on a hollow-core anti-resonant fiber (HC-ARF) was reported for the first time. The HC-ARF was applied as a light transmission medium and gas chamber. The constructed all-fiber structure has merits of low loss, easy optical alignment, good system stability, reduced sensor size and cost. The mid-infrared transmission structure can be utilized to target the strongest gas absorption lines. The reversely-tapered SM1950 fiber and the HC-ARF were spatially butt-coupled with a V-shaped groove between the two fibers to facilitate gas entry. Carbon monoxide (CO) with an absorption line at 4291.50 cm−1 (2.33 µm) was chosen as the target gas to verify the sensing performance. The experimental results showed that the all-fiber LITES sensor based on HC-ARF had an excellent linear response to CO concentration. Allan deviation analysis indicated that the system had excellent long-term stability. A minimum detection limit (MDL) of 3.85 ppm can be obtained when the average time was 100 s