Journal of Shoulder and Elbow Arthroplasty (Feb 2021)

Tomographic Analysis of Positioning of Reverse Baseplates Positioning

  • Alexandre Almeida MBBS,
  • Daniel C Agostini MBBS,
  • Pietro FT Nesello BM,
  • Nayvaldo C de Almeida MBBS,
  • Rafael Mioso MBBS,
  • Ana Paula Agostini MD

DOI
https://doi.org/10.1177/2471549220987714
Journal volume & issue
Vol. 5

Abstract

Read online

Objective To verify whether reverse baseplate positioning without the support of intraoperative three-dimensional technology is within the acceptable parameters in the literature and whether glenoid bone deformity (GBD) compromises this positioning. Methods Sixty-nine reverse shoulder arthroplasties were evaluated with volumetric computed tomography (CT). Two radiologists performed blinded CT scan analysis and evaluated baseplate position within 2mm of the inferior glenoid; the inclination and version of the baseplate in relation to the Friedman line; and upper and lower screw and baseplate metallic peg end point positionings. The patients were divided according to the presence of GBD for statistical analyses. Results The two radiologists concurred reasonably in their interpretations of the following analyzed parameters: baseplate position within 2mm of the inferior glenoid rim (97.1% and 95.7%), baseplate inclination (82.6% and 81.2%), baseplate version (69.6% and 56.5%), the upper screw reaching the base of the coracoid process (71% and 79.7%), the inferior screw remaining inside the scapula (88.4% and 84.1%), and the metallic peg of the baseplate considered intraosseous (88.4% and 72.5%). Conclusion Reverse baseplate positioning without intraoperative three-dimensional technology is within the acceptable parameters of the literature, except for baseplate version and upper screw position. GBD did not interfere with baseplate positioning in reverse shoulder arthroplasty.