Sensors (May 2023)
Temporal, Kinematic and Kinetic Variables Derived from a Wearable 3D Inertial Sensor to Estimate Muscle Power during the 5 Sit to Stand Test in Older Individuals: A Validation Study
Abstract
The 5-Sit-to-stand test (5STS) is widely used to estimate lower limb muscle power (MP). An Inertial Measurement Unit (IMU) could be used to obtain objective, accurate and automatic measures of lower limb MP. In 62 older adults (30 F, 66 ± 6 years) we compared (paired t-test, Pearson’s correlation coefficient, and Bland-Altman analysis) IMU-based estimates of total trial time (totT), mean concentric time (McT), velocity (McV), force (McF), and MP against laboratory equipment (Lab). While significantly different, Lab vs. IMU measures of totT (8.97 ± 2.44 vs. 8.86 ± 2.45 s, p = 0.003), McV (0.35 ± 0.09 vs. 0.27 ± 0.10 m∙s−1, p p p < 0.001) had a very large to extremely large correlation (r = 0.99, r = 0.93, and r = 0.97 r = 0.76 and r = 0.79, respectively, for totT, McT, McF, McV and MP). Bland–Altman analysis showed a small, significant bias and good precision for all the variables, but McT. A sensor-based 5STS evaluation appears to be a promising objective and digitalized measure of MP. This approach could offer a practical alternative to the gold standard methods used to measure MP.
Keywords