Sensors (Jun 2025)
Determination of the Precision of Glucometers Used in Saudi Arabia
Abstract
Background: Efforts have been joined to set the parameters for the reliability of glucometers, yet once they are on the market, they are not further tested for the maintenance of accuracy, specificity, or precision. Methods: This comparative analytical study investigated the precision of commonly used glucometers in Saudi Arabia, namely Accu-Chek Instant®, On-Call Sharp®, and ConTour®, as well as the effects of vitamin C, acetaminophen, and maltose on glucose readings. Ten milliliters of blood was drawn in lithium heparin from healthy volunteers (n = 9). Six samples were divided into two groups of three. One group was designed for normal glucose levels. The second group was designed for high glucose levels by adding a dextrose solution. The last three samples were designed for low glucose levels by leaving the sample for 24 h at room temperature and then following with centrifuge and plasma extraction. Results: This study showed that only Accu-Chek Instant met the International Organization for Standardization (ISO) standard for precision across all dextrose concentrations, along with intra-class correlation values ranging from 0.95–1 (p < 0.001). By spiking the plasma samples with sub-therapeutic, therapeutic, and overdose concentrations of the metabolites, we found that vitamin C had a more evident interference on glucose readings compared to acetaminophen and maltose. Conclusions: The ascertainment of the precision of glucometers and the effects of interferences on them are vital in preventing the improper administration of insulin, which can lead to serious complications.