PLoS ONE (Jan 2021)

Development and ex-vivo assessment of a novel patient specific guide and instrumentation system for minimally invasive total shoulder arthroplasty.

  • Joshua W Giles,
  • Cyrus Brodén,
  • Christine Tempelaere,
  • Roger J H Emery,
  • Ferdinando Rodriguez Y Baena

DOI
https://doi.org/10.1371/journal.pone.0251880
Journal volume & issue
Vol. 16, no. 5
p. e0251880

Abstract

Read online

ObjectiveTo develop and assess a novel guidance technique and instrumentation system for minimally invasive short-stemmed total shoulder arthroplasty that will help to reduce the complications associated with traditional open replacement such as poor muscle healing and neurovascular injury. We have answered key questions about the developed system including (1) can novel patient-specific guides be accurately registered and used within a minimally invasive environment?; (2) can accuracy similar to traditional techniques be achieved?MethodsA novel intra-articular patient-specific guide was developed for use with a new minimally invasive posterior surgical approach that guides bone preparation without requiring muscle resection or joint dislocation. Additionally, a novel set of instruments were developed to enable bone preparation within the minimally invasive environment. The full procedure was evaluated in six cadaveric shoulders, using digitizations to assess accuracy of each step.ResultsPatient-specific guide registration accuracy in 3D translation was 2.2±1.2mm (RMSE±1 SD; p = 0.007) for the humeral component and 2.7±0.7mm (pConclusionThis system achieves minimally invasive shoulder replacement with accuracy similar to traditional open techniques while avoiding common causes of complications.SignificanceThis novel technique could lead to a paradigm shift in shoulder arthroplasty for patients with moderate arthritis, which could significantly improve rehabilitation and functional outcomes.