EPJ Web of Conferences (Jan 2017)

VARIAN CLINAC 6 MeV Photon Spectra Unfolding using a Monte Carlo Meshed Model

  • Morató S.,
  • Juste B.,
  • Miró R.,
  • Verdú G.

DOI
https://doi.org/10.1051/epjconf/201715304012
Journal volume & issue
Vol. 153
p. 04012

Abstract

Read online

Energy spectrum is the best descriptive function to determine photon beam quality of a Medical Linear Accelerator (LinAc). The use of realistic photon spectra in Monte Carlo simulations has a great importance to obtain precise dose calculations in Radiotherapy Treatment Planning (RTP). Reconstruction of photon spectra emitted by medical accelerators from measured depth dose distributions in a water cube is an important tool for commissioning a Monte Carlo treatment planning system. Regarding this, the reconstruction problem is an inverse radiation transport function which is ill conditioned and its solution may become unstable due to small perturbations in the input data. This paper presents a more stable spectral reconstruction method which can be used to provide an independent confirmation of source models for a given machine without any prior knowledge of the spectral distribution. Monte Carlo models used in this work are built with unstructured meshes to simulate with realism the linear accelerator head geometry.