Patient Safety in Surgery (Apr 2023)

Multiphase preclinical assessment of a novel device to locate unintentionally retained surgical sharps: a proof-of-concept study

  • Hae Sung Kang,
  • Jad Khoraki,
  • Jessie Gie,
  • Dielle Duval,
  • Susan Haynes,
  • Michael Siev,
  • Jay Shah,
  • Fernando Kim,
  • Martin Mangino,
  • Levi Procter,
  • Riccardo Autorino,
  • Samuel Weprin

DOI
https://doi.org/10.1186/s13037-023-00359-8
Journal volume & issue
Vol. 17, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Background Retained surgical sharps (RSS) is a “never event” that is preventable but may still occur despite of correct count and negative X-ray. This study assesses the feasibility of a novel device (“Melzi Sharps Finder®” or MSF) in effective detection of RSS. Methods The first study consisted of determination of the presence of RSS or identification of RSS in an ex-vivo model (a container with hay in a laparoscopic trainer box). The second study consisted of determining presence of RSS in an in-vivo model (laparoscopy in live adult Yorkshire pigs) with 3 groups: C-arm, C-arm with MSF and MSF. The third study used similar apparatus though with laparotomy and included 2 groups: manual search and MSF. Results In the first study, the MSF group had a higher rate of identification of a needle and decreased time to locate a needle versus control (98.1% vs. 22.0%, p < 0.001; 1.64 min ± 1.12vs. 3.34 min ± 1.28, p < 0.001). It also had increased accuracy of determining the presence of a needle and decreased time to reach this decision (100% vs. 58.8%, p < 0.001; 1.69 min ± 1.43 vs. 4.89 min ± 0.63, p < 0.001). In-the second study, the accuracy of determining the presence of a needle and time to reach this decision were comparable in each group (88.9% vs. 100% vs. 84.5%, p < 0.49; 2.2 min ± 2.2 vs. 2.7 min ± 2.1vs. 2.8 min ± 1.7, p = 0.68). In the third study, MSF group had higher accuracy in determining the presence of a needle and decreased time to reach this decision than the control (97.0% vs. 46.7%, p < 0.001; 2.0 min ± 1.5 vs. 3.9 min ± 1.4; p < 0.001). Multivariable analysis showed that MSF use was independently associated with an accurate determination of the presence of a needle (OR 12.1, p < 0.001). Conclusions The use of MSF in this study’s RSS models facilitated the determination of presence and localization of RSS as shown by the increased rate of identification of a needle, decreased time to identification and higher accuracy in determining the presence of a needle. This device may be used in conjunction with radiography as it gives live visual and auditory feedback for users during the search for RSS.

Keywords