Zeitschrift für Medizinische Physik (Feb 2023)

EANM guidelines for PET-CT and PET-MR routine quality control

  • Michel Koole,
  • Ian Armstrong,
  • Aron K. Krizsan,
  • Anne Stromvall,
  • Dimitris Visvikis,
  • Bernhard Sattler,
  • Stephan G. Nekolla,
  • John Dickson

Journal volume & issue
Vol. 33, no. 1
pp. 103 – 113

Abstract

Read online

We present guidelines by the European Association of Nuclear Medicine (EANM) for routine quality control (QC) of PET-CT and PET-MR systems. These guidelines are partially based on the current EANM guidelines for routine quality control of Nuclear Medicine instrumentation but focus more on the inherent multimodal aspect of the current, state-of-the-art PET-CT and PET-MR scanners. We briefly discuss the regulatory context put forward by the International Electrotechnical Commission (IEC) and European Commission (EC) and consider relevant guidelines and recommendations by other societies and professional organizations. As such, a comprehensive overview of recommended quality control procedures is provided to ensure the optimal operational status of a PET system, integrated with either a CT or MR system. In doing so, we also discuss the rationale of the different tests, advice on the frequency of each test and present the relevant MR and CT tests for an integrated system. In addition, we recommend a scheme of preventive actions to avoid QC tests from drifting out of the predefined range of acceptable performance values such that an optimal performance of the PET system is maintained for routine clinical use.

Keywords