Journal of Medical Physics (Jan 2016)

Dosimetric evaluation of newly developed well-type ionization chamber for use in the calibration of brachytherapy sources

  • Sathiyan Saminathan,
  • Henry Finlay Godson,
  • Retna Ponmalar,
  • Ravikumar Manickam,
  • James Mazarello

DOI
https://doi.org/10.4103/0971-6203.195187
Journal volume & issue
Vol. 41, no. 4
pp. 234 – 239

Abstract

Read online

The well-type ionization chamber has been designed for convenient use in brachytherapy source strength calibration. The chamber has a volume of 240 cm 3 , weight of 2.5 kg, and is open to atmospheric conditions. The well-type ionization chamber dosimetric characteristics such as leakage current, stability, scattering effect, ion collection efficiency, and nominal response with energy were studied. The evaluated dosimetric characteristics of well-type ionization chamber were compared with two other commercially available well-type ionization chambers. The study shows that the newly developed well-type ionization chamber is reliable for air-kerma strength calibration. The results obtained confirm that this chamber can be used for the calibrations of high-dose rate brachytherapy sources.

Keywords