BMC Gastroenterology (Feb 2024)

Possible use of 2D shear wave liver elastography in new-onset ascites evaluation

  • Andrej Hari,
  • Borut Štabuc

DOI
https://doi.org/10.1186/s12876-024-03159-1
Journal volume & issue
Vol. 24, no. 1
pp. 1 – 9

Abstract

Read online

Abstract Background No data on the use of 2D shear wave elastography exists regarding the evaluation of the new-onset ascites causality. Aims To determine whether 2D shear wave elastography can help in the non-invasive assessment of the new-onset ascites cause. To assess the applicability of liver stiffness measured by 2D shear wave elastography using Esaote MyLab Nine apparatus in patients with ascites. Methods In 52 consecutive patients with new-onset ascites (January 2020 to October 2021), liver stiffness using 2D shear wave elastography was prospectively measured. The reliable measurements were used for further analysis. Relevant clinical and laboratory data was collected. Results The calculated liver stiffness measurement cut-off value of 14.4 kPa held 94% accuracy, 100% sensitivity, and 83% specificity when determining ascites with serum ascites albumin gradient ≥11 g/L. Reliable 2D shear wave elastography success rate was 84%. Conclusions 2D shear wave elastography may potentially be used to differentiate transudative from exudative ascites, especially in patients with portal hypertension and peritoneal carcinomatosis.