Archives of Metallurgy and Materials (Jun 2016)

Potentiostatic, Potentiodynamic and Impedance Study of TiO2 Layers Deposited of 316 LVM Steel Used for Coronary Stents

  • Basiaga M.,
  • Staszuk M.,
  • Walke W.,
  • Tański T.,
  • Kajzer W.

DOI
https://doi.org/10.1515/amm-2016-0138
Journal volume & issue
Vol. 61, no. 2
pp. 821 – 824

Abstract

Read online

The objective of the study is assessment of suitability of the ALD method for application of a TiO2 layer on surface of 316LVM steel used for production of vessel stents. Selection of the appropriate process parameters for application of the layer affects its electrochemical properties, which largely determine resistance of the biomaterial to corrosion in the blood environment, thus affecting its hemocompatibility. To assess resistance of the AISI 316LVM steel with modified surface to corrosion, voltammetric and impedance measurements were conducted. The proposed variant of surface processing allows safe deformation of the TiO2 layer without its delamination.

Keywords