Journal of Cardiovascular Development and Disease (May 2021)

Ablation of Atrioventricular Nodal Re-Entrant Tachycardia Combining Irrigated Flexible-Tip Catheters and Three-Dimensional Electroanatomic Mapping: Long-Term Outcomes

  • Michele Malagù,
  • Francesco Vitali,
  • Federico Marchini,
  • Alessio Fiorio,
  • Paolo Sirugo,
  • Daniela Mele,
  • Alessandro Brieda,
  • Cristina Balla,
  • Matteo Bertini

DOI
https://doi.org/10.3390/jcdd8060061
Journal volume & issue
Vol. 8, no. 6
p. 61

Abstract

Read online

Background: Transcatheter ablation is the standasrd treatment for atrioventricular nodal re-entrant tachycardia (AVNRT). However, different techniques are available. Data about the use of irrigated flexible-tip catheters and three-dimensional electroanatomical mapping (3D EAM) for AVNRT ablation are scant. The aim of this study was to evaluate in long-term follow-up efficacy and safety of a novel approach for AVNRT treatment. Methods: This is a cohort single arm study with long-term follow-up. Patients with AVNRT were treated with catheter ablation by means of irrigated flexible-tip catheters combined with 3D EAM. Results: One-hundred-and-fifty patients were enrolled and followed-up for a median of 38 months (minimum 12, maximum 74). Acute procedural success rate was 96.7% (145/150 patients). During follow-up, 11 patients had arrhythmia recurrences (7.3%). No patient developed atrioventricular conduction block with need for pacemaker implantation (0%). Fourteen patients died during follow-up (9.3%). Conclusions: Acute procedural success and long-term follow-up show that AVNRT could be safely and effectively treated with irrigated flexible-tip catheters and 3D EAM.

Keywords