Sensors (Sep 2019)

A New Configurable Wireless Sensor System for Biomedical Applications with ISO 18000-3 Interface in 0.35 µm CMOS

  • Tatjana Fedtschenko,
  • Alexander Utz,
  • Alexander Stanitzki,
  • Andreas Hennig,
  • Andre Lüdecke,
  • Norbert Haas,
  • Rainer Kokozinski

DOI
https://doi.org/10.3390/s19194110
Journal volume & issue
Vol. 19, no. 19
p. 4110

Abstract

Read online

This article presents a new configurable wireless sensor system. The system is used to perform amperometric measurements and send the measurement data to a handheld reader using a wireless transponder interface. The two-chip sensor system was implemented in a 0.35 μm CMOS technology. The system consists of an integrated nano-potentiostat that performs the actual measurements and an ISO 18000-3 compliant frontend that enables wireless telemetric data transmission and powering of the entire sensor system. The system was manufactured in combination with a chronoamperometric glucose sensor which allows the measurement of the glucose content in tear fluid and thus a non-invasive determination of the blood sugar level. For a range of sensor currents from 0.1 μA to 10 μA, the potentiostat achieved an accuracy of better than 5 % with a total power dissipation of less than 600 μW. With the realized antenna geometry a wireless communication distance of more than 7 cm has been achieved.

Keywords