Assessing Isometric Quadriceps and Hamstring Strength in Young Men and Women: Between-Session Reliability and Concurrent Validity
Abstract
This study assessed the between-session reliability of the FysioMeter H-station when measuring isometric quadriceps and hamstrings strength and H/Q (hamstring/quadriceps) ratio, and the concurrent validity when compared to an isokinetic dynamometer (ID). Sixteen young males and three females (25.6 ± 1.7 years old) performed maximum quadriceps and hamstring contractions using the H-station and the Humac NORM ID over two sessions. Between-session reliability was evaluated by comparing scores from the first and the second session. Concurrent validity was assessed by comparing scores from the H-station with the Humac NORM ID. The H-station showed excellent relative reliability for the quadriceps, hamstring, and the H/Q ratio with ICC2.1 ranging from 0.89 to 0.91. The relative reliability of the Humac NORM was good to moderate for the quadriceps, hamstring, and the H/Q ratio with ICC2.1 ranging from 0.89 to 0.91. Acceptable absolute reliability of the H-station was found across quadriceps, hamstring, and H/Q measures (coefficient of variance (CV) = 7.2 to 9.3%, limit of agreement (LOA) = 20.0 to 24.3%). The validity of the H-station was large for hamstring strength (r = 0.79) and moderate for quadriceps strength (r = 0.69) and H/Q ratio (r = 0.39) compared to the Humac NORM ID. Bland–Altman plots showed LOAs ranging from 37.9 to 59.5%. The results indicate that the H-station can be used to make reliable assessments of relative changes in maximum isometric quadriceps and hamstring strength.