Journal of Orthopaedic Surgery and Research (May 2019)
The 10-year outcomes of the ASR XL Acetabular System: a single-center experience from China
Abstract
Abstract Background The revision rate of articular surface replacement (ASR) implants continues to rise in China because of metal debris. However, there are few reports on the clinical results of ASR implants with prolonged follow-up time in China. This study investigated the clinical outcomes and the risk factors of revision surgery in patients with ASR implants. Methods In total, 74 patients (74 hips) who underwent primary total hip arthroplasty (THA) with ASR implants over the past 4 to 10 years were retrospectively analyzed. Relevant clinical, radiographic, and biochemical data were examined. Results The average follow-up time was 88.46 (range 23–114) months, and the ASR implants of 18 hips (24.3%) were revised. Patients who received revision surgery had worse joint function with significantly lower Harris Hip Score and Western Ontario and McMaster Universities index than non-revision patients (61.11 ± 6.68 vs 85.30 ± 9.16, p 47.80°, HR = 70.145, 95% CI 1.558–3158.213, p = 0.029), high serum Cr concentration (98.44 μg/L, HR = 58.956, 95% CI 1.294–2685.203, p = 0.036), and high serum Co concentration (> 6.95 μg/L, HR = 179.511, 95% CI 2.360–13656.941, p = 0.019) decreased significantly than the lower groups. Conclusions Evaluation of the DePuy ASR XL articulation demonstrated increased rates of revision following a longer follow-up period. High acetabular abduction angle and serum Cr and Co concentration correlated with worse clinical outcomes and high revision rate. Therefore, we advocate that patients with DePuy ASR XL implants be followed up more closely than those with other implants, especially with high acetabular abduction angle and serum Cr or Co concentration.
Keywords