Scientific Reports (Jan 2024)

Measurement accuracy of the acetabular cup position using an inertial portable hip navigation system with patients in the lateral decubitus position

  • Hiromasa Tanino,
  • Ryo Mitsutake,
  • Hiroshi Ito

DOI
https://doi.org/10.1038/s41598-024-51785-2
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Accurate cup placement is critical to ensure satisfactory outcomes after total hip arthroplasty. Portable hip navigation systems are novel intraoperative guidance tools that achieve accurate cup placement in the supine position; however, accuracy in the lateral decubitus position is under debate. A new inertial portable navigation system has recently become available. The present study investigated the accuracy of measurements of the cup position in 54 patients in the lateral decubitus position using this system and compared it with that by a goniometer. After cup placement, cup abduction and anteversion were measured using the system and by the goniometer, and were then compared with postoperatively measured angles. Absolute measurement errors with the system were 2.8° ± 2.6° for cup abduction and 3.9° ± 2.9° for anteversion. The system achieved 98 and 96% measurement accuracies within 10° for cup abduction and anteversion, respectively. The system was more accurate than the goniometer for cup anteversion (p < 0.001), but not for abduction (p = 0.537). The system uses a new registration method of the pelvic reference plane and corrects intraoperative pelvic motion errors, which may affect measurement accuracy. In the present study, reliable and reproducible intraoperative measurements of the cup position were obtained using the inertial portable navigation system.