Non-enzymatic electrochemical sensing platform based on metal oxide-loaded biopolymer for voltammetric measurement of hepatoprotective metadoxine drug in pharmaceutical formulation and human blood serum

  • Manish S. Sengar,
  • Priya Kumari,
  • Neha Sengar,
  • Soami P. Satsangee,
  • Rajeev Jain

Journal volume & issue
Vol. 17
p. 100453

Abstract

Read online

This study focuses on the first-ever high-sensitivity, low-cost, and quick response electrochemical estimation of metadoxine (MTD) in pharmaceuticals and human blood serum using a non-enzymatic nanocomposite modified glassy carbon electrode (CuO/CH/GCE) loaded with chitosan. The electroactive surface of GCE was produced by drop-casting a suspension of CuO/CH nanocomposite in N,N-dimethylformamide (DMF) onto the non-enzymatic electrode surface. Synthesized nanocomposite was characterised by using XRD, XPS, EDX, TEM, Raman and FESEM techniques. EIS technique was utilized to study the enhanced charge-transfer phenomenon occurring at the surface of modified sensor. The electrooxidation of MTD at CuO/CH/GCE surface is depending at pH of supporting electrolyte, scan rate and concentration of analyte. CV and SWV technique were used to carry out electrochemical study, optimised voltammetric response is observed in the BR buffer at pH 2.5, with irreversible diffusion-controlled process. Within the linear concentration range of MTD from 1.99 μg/L to 29.56 μg/L, this sensor exhibited lowest LOD (0.64 μg/L) and LOQ (2.14 μg/L). The MTD in pharmaceutical formulation and human blood serum can be determined with this highly selective peak potential (Ep ∼ 1.2 V), repeatable (%RSD 0.91), and reproducible (%RSD 2.31) method. The average percentage recovery in human blood serum and pharmaceutical formulation is 99.89% and 99.90%. This paper reports a lowest LOD value for MTD detection in the comparison with other reported methods (Table S1). None of the selected excipients were found to interfere more than 5% with redox potential of MTD without affecting the performance of sensor.

Keywords