Antimicrobial Resistance and Infection Control (May 2021)

SARS-CoV-2 aerosol generation during respiratory equipment reprocessing

  • Camila Quartim de Moraes Bruna,
  • Caroline Lopes Ciofi-Silva,
  • Anderson Vicente de Paula,
  • Lucy Santos Villas Boas,
  • Noely Evangelista Ferreira,
  • Tania R. Tozetto-Mendoza,
  • Maria Cássia Mendes Correa,
  • Kazuko Uchikawa Graziano

DOI
https://doi.org/10.1186/s13756-021-00955-2
Journal volume & issue
Vol. 10, no. 1
pp. 1 – 4

Abstract

Read online

Abstract Aerosolization may occur during reprocessing of medical devices. With the current coronavirus disease 2019 pandemic, it is important to understand the necessity of using respirators in the cleaning area of the sterile processing department. To evaluate the presence of severe acute respiratory syndrome coronavirus (SARS-CoV-2) in the air of the sterile processing department during the reprocessing of contaminated medical devices. Air and surface samples were collected from the sterile processing department of two teaching tertiary hospitals during the reprocessing of respiratory equipment used in patients diagnosed with coronavirus disease 2019 and from intensive care units during treatment of these patients. SARS-CoV-2 was detected only in 1 air sample before the beginning of decontamination process. Viable severe acute respiratory syndrome coronavirus 2 RNA was not detected in any sample collected from around symptomatic patients or in sterile processing department samples. The cleaning of respiratory equipment does not cause aerosolization of SARS-CoV-2. We believe that the use of medical masks is sufficient while reprocessing medical devices during the coronavirus disease 2019 pandemic.