Nuclear Technology and Radiation Protection (Jan 2019)
Response of GAFChromic® HD-V2 film dosimeter in 10-300 Gy dose range for radiation testing of electronic devices
Abstract
The study reported in this paper aimed to investigate the response of a GAFChromic® HD-V2 film dosimeter in the dose range between 10 and 300 Gy, normally used to perform irradiation tests on electronic devices with a 60Co gamma-ray irradiator. The well-defined linearity of response in terms of absorbance as a function of absorbed dose, easiness of handling and data analysis of irradiated dosimeters, reproducibility, stability, and insensitivity to visible light and most of the environmental parameters, make HD-V2 film a flexible, inexpensive and reliable dose measurement device. The study has shown a fogging effect of the dosimeter response since its value changes over time. Strategies and adequate precautions to take into account changes on HD-V2 film physical properties for calibration and routine use in industrial applications, with a specific reference to radiation testing of electronic devices, have been discussed.
Keywords