Jornal de Pediatria (Aug 2005)

Acurácia, utilidade e complicações da monitorização subcutânea contínua da glicose (CGMS) em pacientes pediátricos com diabetes tipo 1 Accuracy, utility and complications of continuous glucose monitoring system (CGMS) in pediatric patients with type 1 diabetes

  • Frederico F. R. Maia,
  • Levimar R. Araújo

DOI
https://doi.org/10.1590/S0021-75572005000500006
Journal volume & issue
Vol. 81, no. 4
pp. 293 – 297

Abstract

Read online

OBJETIVO: Avaliar a acurácia, utilidade e complicações da monitorização subcutânea contínua da glicemia em crianças e adolescentes com diabetes melito tipo 1 (DM1). MÉTODOS: Foram estudados retrospectivamente 16 pacientes (16,12±4,41 anos), submetidos à monitorização subcutânea contínua da glicemia (Medtronic; Northridge, CA, EUA) por 72 horas. Foram analisados os valores de glicemia capilar média e pelo sensor monitorização subcutânea contínua da glicemia; excursões glicêmicas (monitorização subcutânea contínua da glicemia versus. glicemia capilar); hiperglicemia pós-prandial (OBJECTIVE: To evaluate the accuracy, utility and complications of continuous glucose monitoring system in children and adolescents with type 1 diabetes. METHODS: This retrospective study assessed 16 type 1 diabetic patients (16.12±4.41 years) submitted to continuous glucose monitoring system (Medtronic; Northridge, CA) for 72 hours. The following parameters were analyzed: mean capillary glucose level and mean glucose value measured by the continuous glucose monitoring system; glucose excursions (continuous glucose monitoring system vs. capillary glucose measurement), postprandial hyperglycemia (NR < 140 mg/dl), nocturnal hypoglycemia, complications (trauma, local infection, disconnection) and therapeutic management after continuous glucose monitoring. A1c levels were measured at the beginning and after 3 months of the study. RESULTS: The mean capillary glucose values were 214.3±66.5 mg/dl vs. 207.6±54.6 mg/dl by continuous glucose monitoring system, with a significant correlation (p = 0.001). The correlation coefficient and mean absolute error were 0.86±0.21 and 12.6% of the median, respectively. The continuous glucose monitoring system was significantly more efficient in detecting glucose excursion than fingerstick capillary blood sampling (p = 0.04; W = 74), and postprandial hyperglycemia was identified in 60% of type 1 diabetic patients with a median value of 157 mg/dl (< 140 mg/dl). Nocturnal hypoglycemia was detected in 46.7% of these patients. The evaluation of A1c levels in eight (50%) patients before continuous glucose monitoring and after 3 months showed a significantly lower level of A1c in this population (8.18±1.5 vs. 7.28±1.3; p = 0.034). The therapeutic management of type 1 diabetes was changed in 100% of patients. No complications were detected in 93.7% of patients. CONCLUSIONS: The continuous glucose monitoring system showed to be a very safe, well-tolerated and highly accurate method, with a low complication rate. It is a good method to identify glucose excursion and postprandial hyperglycemia, and to improve metabolic changes in therapeutic strategies, with a significant impact on the A1c levels of pediatric diabetic patients. The efficacy of the continuous glucose monitoring system in detecting hypoglycemia is still unclear in the medical literature.

Keywords