Ultrasonography (Jul 2020)

Comparison of point and 2-dimensional shear wave elastography for the evaluation of liver fibrosis

  • Sang Min Lee,
  • Min-Jeong Kim,
  • Jeong Hee Yoon,
  • Wonju Hong,
  • Hong Il Ha,
  • Kwanseop Lee,
  • Ji-Young Choe,
  • Jung Woo Lee,
  • Sam-Youl Yoon,
  • Junhee Han

DOI
https://doi.org/10.14366/usg.19090
Journal volume & issue
Vol. 39, no. 3
pp. 288 – 297

Abstract

Read online

Purpose This study aimed to assess the technical performance of ElastQ Imaging compared with ElastPQ and to investigate the correlation between liver stiffness (LS) values obtained using these two techniques. Methods This retrospective study included 249 patients who underwent LS measurements using both ElastPQ and ElastQ Imaging equipped on the same machine. The applicability, repeatability (coefficient of variation [CV]), acquisition time, and LS values were compared using the chi-square or Wilcoxon signed-rank tests. In the development group, the correlation between the LS values obtained by the two techniques was assessed with Spearman correlation coefficients and linear regression analysis. In the validation group, the agreement between the estimated and real LS values was evaluated using a Bland-Altman plot. Results ElastQ Imaging had higher applicability (94.0% vs. 78.3%, P<0.001) and higher repeatability, with a lower median CV (0.127 vs. 0.164, P<0.001) than did ElastPQ. The median acquisition time of ElastQ Imaging was significantly shorter than that of ElastPQ (45.5 seconds vs. 96.5 seconds, P<0.001). The median LS value obtained using ElastQ Imaging was significantly higher than that obtained using ElastPQ (5.60 kPa vs. 5.23 kPa, P<0.001). The LS values between the two techniques exhibited a strong positive correlation (r=0.851, P<0.001) in the development group. The mean difference and 95% limits of agreement were 0.0 kPa (-3.9 to 3.9 kPa) in the validation group. Conclusion ElastQ Imaging may be more reliable and faster than ElastPQ, with strongly correlated LS measurements.

Keywords