Frontiers in Oncology (Nov 2023)

Motion and dosimetric criteria for selecting gating technique for apical lung lesions in magnetic resonance guided radiotherapy

  • Matteo Galetto,
  • Matteo Galetto,
  • Matteo Nardini,
  • Amedeo Capotosti,
  • Guenda Meffe,
  • Davide Cusumano,
  • Luca Boldrini,
  • Giuditta Chiloiro,
  • Angela Romano,
  • Claudio Votta,
  • Maria A. Gambacorta,
  • Maria A. Gambacorta,
  • Luca Indovina,
  • Lorenzo Placidi

DOI
https://doi.org/10.3389/fonc.2023.1280845
Journal volume & issue
Vol. 13

Abstract

Read online

IntroductionPatients treatment compliance increases during free-breathing (FB) treatment, taking generally less time and fatigue with respect to deep inspiration breath-hold (DIBH). This study quantifies the gross target volume (GTV) motion on cine-MRI of apical lung lesions undergoing a SBRT in a MR-Linac and supports the patient specific treatment gating pre-selection.Material and methodsA total of 12 patients were retrospectively enrolled in this study. During simulation and treatment fractions, sagittal 0.35 T cine-MRI allows real-time GTV motion tracking. Cine-MRI has been exported, and an in-house developed MATLAB script performed image segmentation for measuring GTV centroid position on cine-MRI frames. Motion measurements were performed during the deep inspiration phase of DIBH patient and during all the session for FB patient. Treatment plans of FB patients were reoptimized using the same cost function, choosing the 3 mm GTV-PTV margin used for DIBH patients instead of the original 5 mm margin, comparing GTV and OARs DVH for the different TP.ResultsGTV centroid motion is <2.2 mm in the antero-posterior and cranio-caudal direction in DIBH. For FB patients, GTV motion is lower than 1.7 mm, and motion during the treatment was always in agreement with the one measured during the simulation. No differences have been observed in GTV coverage between the TP with 3-mm and 5-mm margins. Using a 3-mm margin, the mean reduction in the chest wall and trachea–bronchus Dmax was 2.5 Gy and 3.0 Gy, respectively, and a reduction of 1.0 Gy, 0.6 Gy, and 2.3% in Dmax, Dmean, and V5Gy, respectively, of the homolateral lung and 1.7 Gy in the contralateral lung Dmax.DiscussionsCine-MRI allows to select FB lung patients when GTV motion is <2 mm. The use of narrower PTV margins reduces OARs dose and maintains target coverage.

Keywords