Applied Sciences (Sep 2021)
The Influence of Laboratory Scanner Versus Intra-Oral Scanner on Determining the Implant Axis by Using Three Different Scan Abutments
Abstract
Background: The purpose of this in vitro study was to compare the implant axis’ spatial position and orientation by using laboratory scanner versus intra-oral scanner with three different scan abutments. Methods: A 3D model was printed with an internal hex implant analog in the place of teeth 35#. Three standard scan abutments were used: MIS (two-piece titanium), AB (two-piece PEEK and titanium base) and ZZ (one-piece PEEK). Each scan abutment was scanned 30 times by TRIOS E3 (laboratory scanner) and 30 times by Omnicam (intra-oral scanner). For each scan, an STL (stereolithography) file was created, and the spatial characterization of each scan abutment was measured in the X, Y, Z coordinates, and rotational and longitudinal angles. The comparison between all the scans was conducted by superimposition of the STL files, using a 3D software. A t-test and Wilcoxon signed-rank test were used. (p p < 0.05). All other scan abutments showed a statistical difference in all axes. The rotational angle of the AB scan abutment was twice the angle of the MIS and ZZ scan abutments. Conclusions: All three scan abutments showed a rotational deviation of the implant axis between the laboratory scanner and the intra-oral scanner. The AB scan abutment showed the greatest deviation (1.04 degrees) while the other two abutments showed deviations of about half a degree in relation to the laboratory scan abutment. There is a need for further studies which will examine the influence of geometry, material, and scan abutment parts on the accuracy of the scan obtained.