BMC Anesthesiology (Jul 2018)

Heating capabilities of small fluid warming systems

  • Norbert Zoremba,
  • Christian Bruells,
  • Rolf Rossaint,
  • Thomas Breuer

DOI
https://doi.org/10.1186/s12871-018-0565-x
Journal volume & issue
Vol. 18, no. 1
pp. 1 – 4

Abstract

Read online

Abstract Background Perioperative temperature management is fundamental to ensure normothermia in patients. Fluid warmers, which have become smaller in size over the past few years, can help to maintain a stable body temperature. Potentially, the reduction of the size may influence the heating performance. Methods Therefore, we tested the effectiveness of enFlow®, Fluido compact® and Thermosens® fluid warmers by measuring the inlet and outlet temperature for room-tempered and ice-cooled saline at flow rates of 25, 50, 75 and 100 ml/min. Results At all examined flow rates, the tested heating devices warmed up room-tempered saline effectively. The enFlow® provided the significantly (p 38 °C at all tested flow rates. The Fluido compact® ensured this only at flow rates of 25 and 50 ml/min, while the Thermosens® provided these conditions at flow rates of 25, 50 and 75 ml/min. Conclusions The heating capability for room-tempered saline was effective in all tested devices, but with ice-cooled saline enFlow® is superior at high flow rates. At low flow rates the heating capabilities of enFlow®, Fluido compact® and Thermosens® are comparable.

Keywords