Sensors (Sep 2013)

Temperature Stability of the Sky Quality Meter

  • Christopher C.M. Kyba,
  • Franz Hölker,
  • Jürgen Fischer,
  • Sabrina Schnitt,
  • Thomas Ruhtz

DOI
https://doi.org/10.3390/s130912166
Journal volume & issue
Vol. 13, no. 9
pp. 12166 – 12174

Abstract

Read online

The stability of radiance measurements taken by the Sky Quality Meter (SQM)was tested under rapidly changing temperature conditions during exposure to a stable lightfield in the laboratory. The reported radiance was found to be negatively correlated withtemperature, but remained within 7% of the initial reported radiance over a temperaturerange of -15 °C to 35 °C, and during temperature changes of -33 °C/h and +70 °C/h.This is smaller than the manufacturer’s quoted unit-to-unit systematic uncertainty of 10%,indicating that the temperature compensation of the SQM is adequate under expected outdoor operating conditions.

Keywords