Sensors (Sep 2013)
Temperature Stability of the Sky Quality Meter
Abstract
The stability of radiance measurements taken by the Sky Quality Meter (SQM)was tested under rapidly changing temperature conditions during exposure to a stable lightfield in the laboratory. The reported radiance was found to be negatively correlated withtemperature, but remained within 7% of the initial reported radiance over a temperaturerange of -15 °C to 35 °C, and during temperature changes of -33 °C/h and +70 °C/h.This is smaller than the manufacturer’s quoted unit-to-unit systematic uncertainty of 10%,indicating that the temperature compensation of the SQM is adequate under expected outdoor operating conditions.