Results of an intercomparison for electric field strength measurements within the German calibration service

  • R. Pape,
  • U. Karsten,
  • F.-M. Lindner,
  • F. Rittmann,
  • J. von Freeden,
  • T. Kleine-Ostmann,
  • T. Schrader

DOI
https://doi.org/10.5194/ars-15-243-2017
Journal volume & issue
Vol. 15
pp. 243 – 248

Abstract

Read online

In this paper we discuss the results of an intercomparison for electric field strength measurements within the German Calibration Service (Deutscher Kalibrierdienst – DKD). The comparison has been carried out on the field strength value required to reach a display reading of 20 V m−1 of the field probes for frequencies between 100 MHz and 18 GHz. Five laboratories joined the intercomparison including the Physikalisch-Technische Bundesanstalt (PTB), the German National Metrology Institute that keeps the primary standard for electric field strength. As measurement artefacts both a small 1-axis probe usually used as transfer sensor at PTB and a larger 3-axis commercial field probe have been used. While the results agree well for the small field probe and when the larger commercial 3-axis field probe is oriented in the direction of the magnetic field, larger deviations occur, when the larger 3-axis field probe is oriented into the direction of the Poynting vector of the calibration field.