PLoS ONE (Jan 2019)
Systematic review and meta-analysis comparing Adjustable Transobturator Male System (ATOMS) and Adjustable Continence Therapy (ProACT) for male stress incontinence.
Abstract
Background and purposeUrinary incontinence is one of the most serious complications of prostate cancer treatment. The objective of this study was to assess efficacy and safety of Adjustable Transobturator Male System (ATOMS) compared to Adjustable Continence Therapy (proACT) for male stress urinary incotinence according to literature findings.Material and methodsA systematic review and meta-analysis on adjustable devices ATOMS and ProACT is presented. Studies on female or neurogenic incontinence were excluded. Differences between ATOMS and proACT in primary objective: dryness status (no-pad or one safety pad/day) after initial device adjustment, and in secondary objectives: improvement, satisfaction, complications and device durability, were estimated using random-effect model. Statistical heterogeneity among studies included in the meta-analysis was assessed using tau2, Higgins´s I2 statistics and Cochran´s Q test.ResultsCombined data of 41 observational studies with 3059 patients showed higher dryness (68 vs. 55%; p = .01) and improvement (91 vs. 80%; p = .007) rate for ATOMS than ProACT. Mean pad-count (-4 vs. -2.5 pads/day; p = .005) and pad-test decrease (-425.7 vs. -211.4 cc; p ConclusionsDespite the limitations that studies available are exclusively descriptive and the follow-up is limited, literature findings confirm ATOMS is more efficacious, with higher patient satisfaction and better durability than ProACT to treat male stress incontinence.